Ethylene oxide sterilizers
They are specially designed for sterilization using pure ethylene oxide or an explosion-proof mixture of ethylene oxide and an inert gas.
Constructed entirely of 316L stainless steel and designed according to the ASME code, it features interior surfaces in contact with the steam with sanitary polishing. The entire unit is thermally insulated.
Horizontal or vertical sliding doors with hermetic sealing by an inflatable silicone rubber gasket. The double-door equipment features safety systems that prevent simultaneous opening of both doors.
Microprocessed command with industrial PLC and HMI touch screen, Title 21 CFR Part 11 compliant, which guarantees data traceability. There are different access levels through passwords to set the different parameters for each cycle.
It has an alphanumeric printer that prints the variable data of the process.
The sterilizers work with a superheated water showering system. The system is based on the recirculation of demineralized water, which falls in a constant shower over the material to be sterilized. Simultaneously, pressure compensation is produced in the chamber, preventing deformation of the load. Water recirculation is carried out by means of a stainless steel pump that sends it to a plate heat exchanger. In the same exchanger, the recirculation water is heated by steam and cooled by cold water. A shell and tube (DTS) type exchanger can be offered as an alternative.
Composed of cargo and transport vehicles to accommodate and move the materials to be processed in the autoclave. Each loading system is analyzed and designed according to the different loads to be processed, being an important point in the design of a production autoclave.
In case of working with pure ethylene oxide, the equipment will be constructed in a way to prevent an explosion. The equipment complies with Ex Proof safety standards.
- Scrubber. Allows for the neutralization of ethylene oxide at the equipment outlet immediately after the degassing stage.
- Scales. Ethylene oxide dosing and cross-control system that allows for gas dosing each cycle and control for overconsumption or gas loss.
| Models | Capacity (liters) |
|---|---|
| 662a | 544 liters |
| 882 | 768 liters |
| 992 | 1032 LTS |
| 1120 | 1200 liters |
| 1140 | 2400 liters |
We manufacture special measurement chambers with capacities exceeding 30,000 liters.